Radiation Protection Dosimetry · 2015 · 16 citations · 15 references
MedicineRadiation OncologyRadiation DosimetryRadiation MeasurementTreatment PlanningBrachytherapy DosimetryDose CalculationIn Vivo DosimetryRadiation Dose MeasurementRadiation TherapyTreatment Planning Dose CalculationBrachytherapy Dose CalculationBrachytherapy Dosimetry VerificationCatheter-based BrachytherapyExternal Beam Radiotherapy DosimetryIndependent VerificationRadiation Dose VerificationTreatment Planning DosimetryTreatment Planning System Dose Verification
Routine in vivo dosimetry is well established in external beam radiotherapy; however, it is restricted mainly to detection of gross errors in high-dose-rate (HDR) brachytherapy due to complicated measurements in the field of steep dose gradients in the vicinity of radioactive source and high uncertainties. The results of in vivo dose measurements using TLD 100 mini rods and TLD 'pin worms' in catheter-based HDR brachytherapy are provided in this paper alongside with their comparison with corresponding dose values obtained using calculation algorithm of the treatment planning system. Possibility to perform independent verification of treatment delivery in HDR brachytherapy using TLDs is discussed.
15