Concepedia

Publication | Open Access

Validation of high-performance liquid chromatography (HPLC) method for quantitative analysis of histamine in fish and fishery products

20

Citations

5

References

2016

Year

Abstract

A high-performance liquid chromatography method is described for quantitative determination and validation of histamine in fish and fishery product samples. Histamine is extracted from fish/fishery products by homogenizing with tri-chloro acetic acid, separated with Amberlite CG-50 resin and C18-ODS Hypersil reversed phase column at ambient temperature (25°C). Linear standard curves with high correlation coefficients were obtained. An isocratic elution program was used; the total elution time was 10 min. The method was validated by assessing the following aspects; specificity, repeatability, reproducibility, linearity, recovery, limits of detection, limit of quantification and uncertainty. The validated parameters are in good agreement with method and it is a useful tool for determining histamine in fish and fishery products.

References

YearCitations