Acta Radiologica Diagnosis · 1985 · 45 citations · 11 references
MedicineAnatomyBioinstrumentationGastroenterologyMovement AnalysisGastrointestinal AssessmentPressure DistributionEsophageal Pressure MeasurementPressure Profile AnalysisSensor SitingSphincter Function TestingSphincter Pressure MeasurementUpper Esophageal Sphincter Physiology
A cineradiographic examination of 9 healthy subjects showed asynchronous movement of a manometric device in relation to movement of the upper esophageal sphincter (UES) during deglutition. A single sensor placed within the UES high pressure zone at rest registered pressures existing outside this zone during the major part of deglutition. The manometric device must therefore comprise at least three sensors placed at different levels and about 10 mm apart if consistent pressure recording within this zone is to be achieved. In order to evaluate the pressure profile within the sphincter adequately, cineradiographic determination of the orientation of the measuring device is necessary. This was made possible by insertion of two mutually perpendicular radiopaque indicators into the catheter.
11
The Pharyngoesophageal Closure Mechanism: A Manometric Study
Charles S. Winans · Gastroenterology · 1972 · 174 citations · Full text
Human upper esophageal sphincter. Response to volume, osmotic, and acid stimuli.
Donald C. Gerhardt, Timothy J. Shuck, Ronald A. Bordeaux et al. · PubMed · 1978 · 147 citations
Medicine, Anatomy, Physiology +9
Human upper esophageal sphincter
Donald C. Gerhardt, Timothy J. Shuck, Ronald A. Bordeaux et al. · Gastroenterology · 1978 · 141 citations · Full text
Medicine, Anatomy, Clinical Sciences +11