Optics Express · 2016 · 10 citations · 51 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImagingMedical ImagingRadiographic ImagingSignal Detection TheoryCone-beam Computed Tomography (Cbct)Observer Performance EvaluationTomographic Quantification
We investigate the detection performance of transverse and longitudinal planes for various signal sizes (i.e., 1 mm to 8 mm diameter spheres) in cone beam computed tomography (CBCT) images. CBCT images are generated by computer simulation and images are reconstructed using an FDK algorithm. For each slice direction and signal size, a human observer study is conducted with a signal-known-exactly/background-known-exactly (SKE/BKE) binary detection task. The detection performance of human observers is compared with that of a channelized Hotelling observer (CHO). The detection performance of an ideal linear observer is also calculated using a CHO with Laguerre-Gauss (LG) channels. The detectability of high contrast small signals (i.e., up to 4-mm-diameter spheres) is higher in the longitudinal plane than the transverse plane. It is also shown that CHO performance correlates well with human observer performance in both transverse and longitudinal plane images.
51
L.A. Feldkamp, L. C. Davis, J. W. Kress · Journal of the Optical Society of America A · 1984 · 6.2K citations
Lung Cancer Screening with CT: Mayo Clinic Experience
Stephen J. Swensen, James R. Jett, Thomas E. Hartman et al. · Radiology · 2003 · 676 citations
Evaluation of Patients With Pulmonary Nodules: When Is It Lung Cancer?
Michael K. Gould, James Fletcher, Mark D. Iannettoni et al. · CHEST Journal · 2007 · 509 citations