Experimental determination of dose calibrator settings and study of associated volume dependence in v-vials for rhenium-186 perrhenate solution sources.

14

Citations

6

References

2000

Year

Abstract

These studies indicate a strong volume dependence on the response of the dose calibrator and highlight the need for experimental verification of dose calibrator settings for nonstandard geometries.