Publication | Open Access
Longevity of implantable cardioverter-defibrillators for cardiac resynchronization therapy in current clinical practice: an analysis according to influencing factors, device generation, and manufacturer
44
Citations
18
References
2015
Year
The device longevity at 5 years was 54%. High left ventricular lead output and unipolar pacing configuration were associated with early battery depletion, while recent-generation CRT-Ds displayed better longevity. Significant differences emerged among currently available CRT-D systems from different manufacturers.