Orientation and wear of the acetabular component in third generation alumina-on-alumina ceramic bearings

Patrick James Lusty, Anthony D. Watson, M. Tuke, William L. Walter, William L. Walter, B. Zicat

Journal of Bone and Joint Surgery - British Volume · 2007 · 93 citations · 25 references

DOIFull text

Open access

Concepts

Abstract

We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm(3)/yr for the femoral heads, and 0.04 mm(3)/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of < 15 degrees (seven femoral heads) had a median femoral head wear rate of 1.2 mm(3)/yr, compared with 0 mm(3)/yr for those with an anteversion of > or =15 degrees (15 femoral heads, p < 0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.

References