First measurements with new high-resolution gadolinium-GEM neutron detectors

D. Pfeiffer, F. Resnati, Jens Birch, Maddi Etxegarai, R. Hall-Wilton, Carina Höglund, L. Hultman, Isabel Llamas‐Jansa, E. Oliveri, Esko Oksanen,

Journal of Instrumentation · 2016 · 36 citations · 30 references

DOIFull text

Open access

Concepts

Abstract

European Spallation Source instruments like the macromolecular diffractometer, NMX, require an excellent neutron detection efficiency, high-rate capabilities, time resolution, and an unprecedented spatial resolution in the order of a few hundred micrometers over a wide angular range of the incoming neutrons. For these instruments solid converters in combination with Micro Pattern Gaseous Detectors (MPGDs) are a promising option. A GEM detector with gadolinium converter was tested on a cold neutron beam at the IFE research reactor in Norway. The {\\mu}TPC analysis, proven to improve the spatial resolution in the case of $^{10}$B converters, is extended to gadolinium based detectors. For the first time, a Gd-GEM was successfully operated to detect neutrons with an estimated efficiency of 10% at a wavelength of 2 {\\AA} and a position resolution better than 350 {\\mu}m.

References

30