Publication | Closed Access
Cochlear implant extrusion in a child with keratitis, ichthyosis and deafness syndrome
30
Citations
15
References
1997
Year
We present a child with keratitis, ichthyosis and deafness (KID) syndrome implanted with a Nucleus device. We discuss the would complications of this child and the steps taken to deal with the problems encountered when the wound failed to heal, followed by the partial extrusion of her implant. Early surgical management involved resuturing the wound but when this failed a rotational flap was required to cover the implant package and allow eventual healing. Despite the wound problems and revision surgery she has a good audiological result.