Journal of Bone and Joint Surgery - British Volume · 1997 · 125 citations · 23 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryJoint Replacement SurgeryOrthopaedic BiomaterialsOrthopedicsAseptic LooseningBone CementsCement-retained ProsthesesHip ArthroplastyJoint Replacement MaterialsProsthetic MaterialsFemoral Component DesignFemoral LooseningProsthesis LooseningProsthesis Wear MechanismsTitanium MaterialsTitanium Nitride CoatingsWearWear Of Prosthetic Materials
We have studied aseptic loosening of the femoral component in 76 patients with primary total hip replacement using the Capital prosthesis. The mean follow-up was 26 months (10 to 37). Twelve femoral components (16%) were definitely and eight (10%) were possibly loose. They were characterised by a thin cement mantle (p < 0.001) and excessive residual cancellous bone in the proximomedial region (p < 0.01). We recommend that the cement mantle around the prosthesis should be 2 to 3 mm and that further long-term studies are needed to evaluate the wear properties of titanium-nitride-coated titanium femoral heads.
23