Non-contact measurement of oxygen saturation with an RGB camera

Alessandro Guazzi, Mauricio Villarroel, João Jorge, Jonathan Daly, Matthew Frise, Peter A. Robbins, Lionel Tarassenko

Biomedical Optics Express · 2015 · 148 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

A novel method (Sophia) is presented to track oxygen saturation changes in a controlled environment using an RGB camera placed approximately 1.5 m away from the subject. The method is evaluated on five healthy volunteers (Fitzpatrick skin phenotypes II, III, and IV) whose oxygen saturations were varied between 80% and 100% in a purpose-built chamber over 40 minutes each. The method carefully selects regions of interest (ROI) in the camera image by calculating signal-to-noise ratios for each ROI. This allows it to track changes in oxygen saturation accurately with respect to a conventional pulse oximeter (median coefficient of determination, 0.85).

References

24