Publication | Closed Access
Circuit Impedance Could Be a Crucial Factor Influencing Radiofrequency Ablation Efficacy and Safety: A Myocardial Phantom Study of the Problem and Its Correction
42
Citations
39
References
2015
Year
The lesion and overheated dimensions were significantly larger with lower circuit impedance during irrigated RF ablation and the lesion size was smaller in high impedance ablations. Power delivery adjusted to impedance using a simple equation improved the consistency of lesion formation and prevented overheating.
| Year | Citations | |
|---|---|---|
Page 1