Clinical Chemistry · 1986 · 66 citations · 0 references
Life SciencesBiochemistryClinical BiochemistryClinical ImmunologyGood Laboratory PracticeAntigen Detection AssaysEnzyme-linked Immunosorbent AssayImmunoassay AnalysisImmunoassay DevelopmentAlbumin Analysis In Body FluidsAlbumin QuantificationAntiserum PreparationImmunoassay MethodologyImmunoassaysMidstream Urine CollectionReference Interval DeterminationSerum Creatinine Assessment
We have developed and validated simple, rapid immunoassays to measure concentrations of albumin in urine ranging from 5 to 200 mg/L. We use antiserum to human albumin, raised in sheep, and we separate the antibody-bound and free fractions of albumin by using a second antiserum, produced in donkeys, against the Fc fragment of sheep IgG. These two antisera can be mixed and added to assay tubes as a single reagent without inhibiting antigen binding. Samples or standards are incubated for at least 30 min with fluorescein-labeled albumin and the premixed antiserum reagent. After brief centrifugation, the supernate is discarded and the precipitate containing the bound fraction is dissolved and its fluorescence measured. Alternatively, 125I-labeled albumin can be used as tracer with all other reagents unchanged; in this case one simply counts the radioactivity of the bound fraction. A reference interval of less than 0.2-2.8 mg/mmol of creatinine was obtained for untimed, daytime, midstream urine specimens from 80 healthy subjects.