Orthopedics · 2001 · 11 citations · 17 references
MedicineOrthopaedic SurgerySurgeryBone GraftingJoint Replacement SurgeryOrthopedicsHip ArthroplastyImplant FixationAcetabular Component DesignAcetabular Component FixationAcetabular LooseningAseptic Loosening AssessmentClinical Outcome MeasurementFemoral Component DesignFemoral Component FixationFemoral Stem Design
Outcome of the acetabular component in 90 consecutive primary noncemented total hip arthroplasties (THAs) was prospectively studied. The acetabular cup consisted of a hemispherical titanium alloy shell with a titanium fiber-mesh porous coating and a modular polyethylene liner (Harris-Galante Porous-1, Zimmer, Warsaw, Ind). The cup was implanted using line-to-line reaming with adjunctive dome screw fixation. The femoral component consisted of a titanium alloy stem with titanium fiber-mesh porous coating and a 28-mm cobalt-chrome modular head. Mean patient age was 53 years (range: 27-75 years); male:female ratio was 48:42; and mean follow-up was 6 years (range: 4.5-8 years). One acetabular component was revised for aseptic loosening. Of 81 unrevised hips available for follow-up, mean Harris hip score was 57 preoperatively and 96 at final follow-up (72% excellent, 15% good, 1 3% fair, and none poor). Of 61 unrevised hips with adequate radiographic follow-up, radiographic failure (complete periprosthetic radiolucency) was evident in 3 (4.9%) and periacetabular osteolysis in none. Radiographic failure did not correlate with poor clinical outcome. Linear polyethylene wear rate (mean: 0.13 mm/year) did not correlate with age, gender, weight, outcome, or cup abduction angle, but did correlate with the presence of femoral periprosthetic osteolysis (0.18 mm/year with femoral osteolysis versus 0.11 mm/year without; P= .01). This series of porous-coated hemispherical cups demonstrated excellent intermediate-term clinical and radiographic outcome, comparable with similarly favorable results reported by the prosthesis designers. A potentially adverse effect of polyethylene wear on the longevity of a THA was supported by a positive correlation between polyethylene wear rate and femoral osteolysis.
17