MedicineOtolaryngologyRadiologyDiagnostic ImagingDiagnostic RadiologyOtologyEar AnatomyEar ExaminationMagnetic Resonance Contrast MechanismsMagnetic Resonance ImagingMedical Imaging3D Magnetic Resonance ImagingContrast EnhancementFast Spin EchoImage Acquisition QualityInner Ear AssessmentRadiographic Image QualitySpin Echo Imaging
We compared the image quality of the newly developed ultra-long echo train length (ETL) 3-D fast spin-echo (FSE) and half-Fourier technique, which is performed in less than 3 minutes, with the conventional 3D-FSE imaging technique, which takes 15 minutes, in assessing MR examinations of the inner ear. The new method's images were almost comparable to the conventional 3D-FSE images in depicting anatomic details and pathologic findings. Implementation of the ultra-long ETL and half-Fourier 3D-FSE imaging technique enables acquisition of inner ear MR studies in a vastly reduced time and with high spatial resolution without significant penalty, opening the possibility for low-cost screening of acoustic tumors without contrast enhancement in less than 3 minutes.
3
Steady-state free precession imaging of the inner ear.
Arthur E. Stillman, Kent B. Remley, Daniel J. Loes et al. · PubMed · 1994 · 12 citations · Full text
Medicine, Otolaryngology, Radiology +16