Asymmetric Mach–Zehnder Interferometer Based Biosensors for Aflatoxin M1 Detection

Romain Guider, Manuela Zanetti, Floris Falke, Erik Schreuder, René Heideman, Cecilia Pederzolli, Lorenzo Pavesi

Biosensors · 2016 · 97 citations · 20 references

DOIFull text

Open access

Concepts

Abstract

In this work, we present a study of Aflatoxin M1 detection by photonic biosensors based on Si₃N₄ Asymmetric Mach-Zehnder Interferometer (aMZI) functionalized with antibodies fragments (Fab'). We measured a best volumetric sensitivity of 10⁴ rad/RIU, leading to a Limit of Detection below 5 × 10(-7) RIU. On sensors functionalized with Fab', we performed specific and non-specific sensing measurements at various toxin concentrations. Reproducibility of the measurements and re-usability of the sensor were also investigated.

References