Concepedia

Publication | Closed Access

Centrifugal Analyzer Determination of Ascorbate in Serum or Urine with Fe3+/Ferrozine

24

Citations

0

References

1975

Year

Abstract

Measurement of serum ascorbate may be useful in long-term population studies because of the possible influence of ascorbate on numerous physiological factors. We describe an automated method for determining ascorbate in serum and urine by using the reduction of ferric iron by ascorbate and the formation of a color between the resulting ferrous iron and Ferrozine [3-(2-pyridyl)-5,6-bis(4-phenylsulfonic acid)-1,2,4-triazine]. A centrifugal analyzer is used to rapidly and simultaneously measure ascorbate in the samples and standards and minimize interference from slower reacting substances in the sample. The method is highly precise and specific. Data are also presented on the stability of ascorbate in serum, urine, and aqueous solutions.