Clinical Chemistry · 1973 · 20 citations · 0 references
Life SciencesBiochemistryClinical BiochemistryEnzymologyMetabolismBioanalytical ChemistryBioanalytical Method ValidationBilirubin Analysis In Body FluidsEnzyme ConcentrationGlucose Analysis In Body FluidsGlucose OxidaseProtein Precipitation
Abstract We have modified an ultramicro method for determining glucose with glucose oxidase—peroxidase—odianisidine. Capillary blood is diluted in isotonic NaF (17.6-17.8 g/ liter), the cells are removed, and glucose is determined after a 20-min incubation with enzyme-containing reagent. Increasing the enzyme concentration fourfold nullifies interference from ascorbic acid and hemolysis, while greatly decreasing interference from bilirubinemia. If necessary, bilirubin is removed completely by coprecipitation with protein. Interference from uric acid is minimized by incorporating it into the standards. Several other suspected interferences proved inconsequential. Isotonic NaF does not inhibit the enzymes used, and preserves glucose in blood for at least 2 h. Avoiding protein precipitation and shortening the incubating time significantly lessens analytical time without affecting precision.