Publication | Open Access
Preclinical Assessment of a Fully Automated Multiplex PCR Panel for Detection of Central Nervous System Pathogens
89
Citations
10
References
2015
Year
We evaluated a multiplexed PCR panel for the detection of 16 bacterial, viral, and fungal pathogens in cerebrospinal fluid. Panel results were compared to routine testing, and discrepancies were resolved by additional nucleic acid amplification tests or sequencing. Overall, the positive and negative agreements across methods were 92.9% and 91.9%, respectively.