Clinical Chemistry · 1976 · 75 citations · 0 references
PharmacyPharmaceutical AnalysisPharmacologyChromatographic AnalysisClinical PharmacologyForensic Pharmaceutical AnalysisAnalytical ChromatographyChromatographic Analysis Of PharmaceuticalsChromatographyLiquid ChromatographyAnticonvulsant DrugsAnticonvulsant PharmacologyInternal Standard AnalysisSerum Drug Concentration Measurement
We describe procedures for simultaneously determining some anticonvulsants (phenobarbital, diphenylhydantoin, primidone, ethosuximide, methsuximide, carbamazepine) in serum by high-pressure liquid chromatography. The drugs, together with an internal standard, phenacetin, are adsorbed from serum onto charcoal and eluted from it with organic solvent. The eluate is analyzed isocratically on a reverse-phase column with a mobile phase consisting of acetonitrile/water (17/83 by volume). The eluted drugs are detected by their absorption at 195 nm, and quantities estimated from their peak areas as compared with those of extracted standards. Results are reproducible to about 6%. Sensitivities, for 0.5-ml serum samples, of 0.1 mg/liter for all the drugs analyzed except ethosuximide (0.5 mg/liter) are attained routinely. Correlation of results with gas chromatography was 0.912 for phenobarbital, 0.982 for diphenylhydantoin, 0.886 for primidone, and 0.966 for ethosuximide. Amobarbital and secobarbital interfere with the analysis; chlordiazepoxide, methaqualone, salicylate, diazepam, and oxazepam do not. Including extraction, analysis time for a single sample is 20 min.