2002 · 10 citations · 0 references
EngineeringBioengineeringBiomedical EngineeringBiomedical Signal ProcessingUltrasonographyUltrasoundDigital Signal ProcessingMedical ImagingNoiseNoise ReductionUltrasonicsUltrasound ImagingBackscatter AnalysisData ProcessingFramesImage Noise ReductionImaging Signal ProcessingNoise Reduction AlgorithmsNoise Reduction EvaluationSignal Averaging
Data averaging is a common noise reduction practice in several data processing procedures. The authors have used this approach in order to minimize the noise of the ultrasound signals employed in the backscatter analysis of the heart. They assess the optimum number of frames to be used in the averaging routine in order to obtain, in the search for maximum noise reduction, a good compromise between the two opposite criteria: indicating on the one hand the use of the greatest number of frames and the other hand the smallest. Results obtained show that 10 frames averaging allows 50% noise reduction which is equivalent to 90% of the noise reduction obtained by averaging a 4 times greater number of frames.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>