'Stiction-friction' of total hip prostheses and its relationship to loosening

Sheldon R. Simon, IL Paul, R. M. Rose, Radin El

Journal of Bone and Joint Surgery · 1975 · 39 citations · 0 references

Concepts

Abstract

The static friction, or "stiction-friction", in McKee-Farrar and Charnley-Müller prostheses in a hip joint simulator was compared with dynamic friction determined while the prostheses were oscillated. Under physiological conditions stiction-friction differed little from dynamic friction in both metal-on-metal and metal-on-plastic prostheses, and was affected very little by the lubricant as long as some fluid was present. Stiction-friction increased significantly only after relatively long stationary periods and high loads. However, the frictional forces generated in total hip-replacement prostheses were at least forty times higher than those generated in normal joints and may well be enough to cause late loosening of acetabular components by fatigue failure.