The Long-Term Stability of Portable Spirometers Used in a Multinational Study of the Prevalence of Chronic Obstructive Pulmonary Disease

Rogelio Pérez‐Padilla, Juan Carlos Vázquez-García, María Nelly Márquez, José Roberto Jardim, Julio A. Pertuzé, Carmen Lisboa, Adriana Muiño, María Victorina López, Carlos Tálamo, María Montes de,

Respiratory Care · 2006 · 72 citations · 10 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: We report the performance of an ultrasound-based portable spirometer (EasyOne) used in a population-based survey of the prevalence of chronic obstructive pulmonary disease, conducted in 5 Latin American cities: São Paulo, Brazil; México City, México; Montevideo, Uruguay; Santiago, Chile; and Caracas, Venezuela (the Latin American COPD Prevalence Study [PLATINO]). METHODS: During the survey period (which ranged from 3 months to 6 months in the various locations) we collected daily calibration data from the 70 EasyOne spirometers used in the 5 survey cities. The calibrations were conducted with a 3-L syringe, and the calibration data were stored in the spirometer's database. RESULTS: Ninety-seven percent of the calibration volumes were within +/- 64 mL (2.1%) of the 3-L calibration signal. Excluding data from the first city studied (São Paulo), where one calibration syringe had to be replaced, 98% of the calibration checks were within +/- 50 mL (1.7%). The measured volume was affected only minimally by the syringe's peak flow or emptying time. CONCLUSION: In these 70 EasyOne spirometers neither calibration nor linearity changed during the study. Such calibration stability is a valuable feature in spirometry surveys and in the clinical setting.