Journal of Applied Physiology · 1959 · 36 citations · 0 references
Life SciencesBiochemistryPhysiologyAnalytical ChemistryGood Laboratory PracticeHematology PhysiologyBioanalytical Method ValidationHemostatic AgentsAssessment Of Oxygen SaturationHemolysis AssaysHemolytic Activity TestingOxygen SaturationOxygen Saturation MeasurementSpectroscopic Oxygenation MeasurementStandard Method Of Measurement
A spectrophotometric method for the measurement of blood O 2 saturation utilizing Triton X-100 as the hemolytic agent is described. The results of its comparison with the standard manometric method indicate that it provides reliable values for blood O 2 saturation. Triton X-100 was found to be a more suitable hemolytic agent than saponin. A nomogram is presented which facilitates the calculation of percentage of O 2 saturation from the measurements required by the method. Submitted on August 27, 1958