Journal of Bone and Joint Surgery - British Volume · 2007 · 52 citations · 37 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryOrthopaedic BiomaterialsOrthopedic BiomaterialsOrthopedicsHip ArthroplastyHip DisordersImplant DesignJoint Replacement MaterialsFemoral Stem DesignHighly Cross-linked PolyethyleneHighly Crosslinked PolyethyleneMann - Whitney U TestPenetration RatePolyethylene ImplantsPolyethylene Insert Design
We carried out a prospective randomised study designed to compare the penetration rate of acetabular polyethylene inserts of identical design but different levels of cross-linking at a minimum of four years follow-up. A total of 102 patients (102 hips) were randomised to receive either highly cross-linked Durasul, or contemporary Sulene polyethylene inserts at total hip replacement. A single blinded observer used the Martell system to assess penetration of the femoral head. At a mean follow-up of 4.9 years (4.2 to 6.1) the mean femoral head penetration rate was 0.025 mm/year (SD 0.128) in the Durasul group compared with 0.106 mm/year (SD 0.109) in the Sulene group (Mann-Whitney test, p = 0.0027). The mean volumetric penetration rate was 29.24 mm(3)/year (SD 44.08) in the Durasul group compared with 53.32 mm(3)/year (SD 48.68) in the Sulene group. The yearly volumetric penetration rate was 55% lower in the Durasul group (Mann-Whitney test, p = 0.0058). Longer term results are needed to investigate whether less osteolysis will occur.
37
Orhun K. Muratoglu, C.R. Bragdon, Daniel O’Connor et al. · The Journal of Arthroplasty · 2001 · 640 citations
Natural Sciences, Materials Engineering, Materials Science +16
Three-dimensional polyethylene wear of a press-fit titanium prosthesis
P. Devane, J. G. Horne, Kerryn Martin et al. · The Journal of Arthroplasty · 1997 · 278 citations