Publication | Open Access
Validity of synthetic bone as a substitute for osteoporotic cadaveric femoral heads in mechanical testing
43
Citations
18
References
2012
Year
This study demonstrates that synthetic material with a density of 0.16 g/cm(3) can provide a good substitute for cadaveric osteoporotic femoral heads in the testing of implants. However we do recognise that no synthetic material can be considered as a definitive substitute for bone, therefore studies performed with artificial bone substrates may need to be validated by further testing with a small bone sample in order to produce conclusive results.