2007 · 17 citations · 0 references
EngineeringBioengineeringAnalytical ChemistryBiosensorsSensorsFlow SensorsStreamflow MeasurementDigital Data CollectionField Tracer TestsSensitivityTracer Techniques
In order to simplify sampling dye tracers in water streams without having to invest money in expensive systems, we developed a highly sensitive flow-through field fluorometer. This apparatus was tested successfully in the field with fluorescein and sulforhodamine. A submersible probe is connected to a digital data logger and the data is written to a PCMCIA memory card allowing 2 weeks of unattended recording at 1 sample every 4 minutes. A sensitivity as low as 0.05 ppb has been achieved. Simultaneous concentration measurements of 2 tracers is possible, at the expense of reduced sensitivity, however. Resume