HPLC METHOD FOR DETERMINATION OF LACOSAMIDE S(-)ENANTIOMER IN BULK AND PHARMACEUTICAL FORMULATION

V. Kalyan Chakravarthy, D. Gowri Shankar

2011 · 13 citations · 1 references

Concepts

Abstract

A High performance liquid chromatography (HPLC) method has been developed and subsequently validated for the determination of Lacosamide S (-) Enantiomer in bulk and pharmaceutical formulation. Seperation was achieved in Isocratic mode with a Regis Pack ((Make: Regis Technologies (USA) ; 250 mm x 4.6 mm I.D; particle size 5 µm)) column with mobile phase containing 100% Ethanol as eluent at flow rate 1.0mL/minute and the column temperature was 25°C. UV detection was performed at 210nm. The injection volume is 10 µL. The method is simple, rapid, and selective. The described method of Lacosamide S (-) Enantiomer is linear over a range of 0.1280 mcg/mL to 3.0525 mcg/mL . The method precision for the determination of Lacosamide S (-) Enantiomer was below 2.0%RSD. Limit of Detection (LOD) and Limit of Quantification (LOQ) of Lacosamide S (-) Enantiomer was established and ranged from 0.0427 for LOD and 0.1280 for LOQ . The method is useful in the quality control of Bulk manufacturing and also in pharmaceutical formulations.