Normally Off ECG SoC With Non-Volatile MCU and Noise Tolerant Heartbeat Detector

Shintaro Izumi, Ken Yamashita, Masanao Nakano, Shusuke Yoshimoto, Tomoki Nakagawa, Yozaburo Nakai, Hiroshi Kawaguchi, Hiromitsu Kimura, Kyoji Marumoto, Takaaki Fuchikami,

IEEE Transactions on Biomedical Circuits and Systems · 2015 · 40 citations · 26 references

Concepts

Abstract

This paper describes an electrocardiograph (ECG) monitoring SoC using a non-volatile MCU (NVMCU) and a noise-tolerant instantaneous heartbeat detector. The novelty of this work is the combination of the non-volatile MCU for normally off computing and a noise-tolerant-QRS (heartbeat) detector to achieve both low-power and noise tolerance. To minimize the stand-by current of MCU, a non-volatile flip-flop and a 6T-4C NVRAM are used. Proposed plate-line charge-share and bit-line non-precharge techniques also contribute to mitigate the active power overhead of 6T-4C NVRAM. The proposed accurate heartbeat detector uses coarse-fine autocorrelation and a template matching technique. Accurate heartbeat detection also contributes system-level power reduction because the active ratio of ADC and digital block can be reduced using heartbeat prediction. Measurement results show that the fully integrated ECG-SoC consumes 6.14 μ A including 1.28- μA non-volatile MCU and 0.7- μA heartbeat detector.

References

26