A Comparison of Two Different Systems for CD34+ Selection of Autologous or Allogeneic PBSC Collections

28

Citations

28

References

1998

Year

Abstract

This study compared CD34 selection procedures using the CellPro CEPRATE and the Baxter Isolex 300 systems. Thirty-two procedures were performed, 19 CEPRATE and 13 Isolex. Median starting CD34 percentages were (CEPRATE/Isolex) 0.80% (range 0.24%-7.73%) and 0.85% (range 0.27%-10.17%), respectively (p = 0.788). After selection, there was a highly significant difference in purity of the product (CEPRATE/Isolex), 54% and 82% respectively (p < 0.0001). There was no significant difference in median recovery (CEPRATE/Isolex), 43% and 50%, respectively (p = 0.383). The starting CD34 percentage influenced the purity of the final product, and at high and low starting percentages, the Isolex produced superior purity. Improved efficacy of T cell depletion was observed with the Isolex, a median log depletion of 3.4 compared with 2.9 for the CEPRATE system (p = 0.012). In conclusion, the Isolex 300i produced a significantly higher purity CD34+ fraction, even at starting CD34+ levels of <0.5%, with no significant difference in recovery when compared with the CEPRATE system. The associated log T cell depletion is significantly improved with the Isolex system, with possible implications for use in CD34-selected allogeneic transplants.