IEEE Transactions on Nuclear Science · 1996 · 26 citations · 7 references
EngineeringNuclear EngineeringRadiologyDiagnostic RadiologyNuclear InstrumentationRadiation DetectionGamma-ray DetectionGamma-ray InstrumentationGamma-ray SourcesRadiation Imaging DetectorsRadiation Imaging SystemsRadiographic ImagingAbsolute EfficiencyAngular ResolutionCount Rate AnalysisCount Rate EstimationDetector Spatial ResolutionGamma Camera ImagingGamma-ray ImagingImaging Resolution
A combined video and gamma ray imaging system has been developed which can rapidly determine the location, distribution, and intensity of gamma ray sources. This instrument includes both a conventional video camera and a gamma ray imaging system, which is based upon a position sensitive photomultiplier tube, a scintillator, and a pinhole collimator. The gamma camera records the position and energy of each interaction, determining the energy spectrum and count rate from each direction. We have used a prototype of such an instrument in preliminary field tests to image radioactive sources with gamma ray energies between 120 keV and 2.4 MeV. At 662 keV, this new system achieves an angular resolution of 4/spl deg/ FWHM with an absolute efficiency of 1/spl times/10/sup -6/ at 1 meter, with sensitivity such that a 0.5 mCi /sup 137/Cs source at 5 meters can be located in <60 seconds. Higher efficiency and sensitivity can be achieved at lower spatial resolution.
7
<i>Measurement and Detection of Radiation.</i>
O. Gayou · Medical Physics · 2012 · 559 citations · Full text