IEEE Transactions on Nuclear Science · 2001 · 69 citations · 21 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyHybrid ImagingBreast Cancer ImagingBreast ImagingMammography PhysicsMedical ImagingPlanar ScintigraphyTumor DetectionBreast Phantom StudiesBreast ScintimammographyDual-modality ImagingGamma Camera ImagingGamma ScintigraphyMammographyMammography ImagingPlanar Detectors
Results are presented on the optimization of the design and use of dedicated compact scintimammography gamma cameras. Prototype imagers with a field of view (FOV) of 5/spl times/5 cm/sup 2/, 10/spl times/10 cm/sup 2/ and 15/spl times/20 cm/sup 2/ were used in either a dual modality mode as an adjunct technique to digital X-ray mammography imagers or as stand-alone instruments such as dedicated breast SPECT and planar imagers. Experimental data were acquired to select the best imaging modality (SPECT or planar) to detect small lesions using Tc/sup 99m/ radio-labeled pharmaceuticals. In addition, studies were performed to optimize the imaging geometry. Results suggest that the preferred imaging geometry is planar imaging with two opposing detector heads while the breast is under compression. However, further study of the dedicated breast SPECT is warranted.
21
Frans J.Th. Wackers, Daniel S. Berman, Jamshid Maddahi et al. · PubMed · 1989 · 632 citations
Feasibility study for positron emission mammography
Christopher J. Thompson, K. Murthy, I. Weinberg et al. · Medical Physics · 1994 · 170 citations
Design and evaluation of an LSO PET detector for breast cancer imaging
Niraj K. Doshi, Yiping Shao, Robert W. Silverman et al. · Medical Physics · 2000 · 115 citations
Technetium-99m-sestamibi uptake in breast tumor and associated lymph nodes.
J. Maublant, M de Latour, D. Mestas et al. · PubMed · 1996 · 106 citations