Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: The Effects of Oximeter Probe Type and Gender
373
Citations
4
References
2007
Year
Multivariate analysis indicated that Sao2 level, sensor type, skin color, and gender were predictive of errors in Spo2 estimates at low Sao2 levels. The data suggest that clinically important bias should be considered when monitoring patients with saturations below 80%, especially those with darkly pigmented skin; but further study is needed to confirm these observations in the relevant populations.