Acta Chromatographica · 2013 · 27 citations · 20 references
PharmacyPharmaceutical AnalysisPharmaceuticsChromatographic AnalysisPsychiatric PharmacotherapyPsychopharmacologyAntipsychotic PharmacologyChromatographic Analysis Of PharmaceuticalsDrug EvaluationPharmaceutical AssessmentAtypical Antipsychotic TreatmentChromatographic Drug AnalysisMobile Phase PolarityPurity ControlReagent Purity ControlSolvent Purity Control
Aripiprazole is a novel atypical antipsychotic drug used in the treatment of schizophrenia. The sensitive and reproducible ion pair RPLC method was developed and validated for determination of aripiprazole and its nine impurities, which are significantly different in polarity. The separation was performed on Phenomenex Luna® C18 column (5.0 μm particle size, 250 × 4.6 mm id) using a gradient mobile phase A (phosphate buffer pH 3.0) and mobile phase B (acetonitrile) at the working temperature of 25°C. The buffer was 1.11 g KH2PO4 with 1.2 g sodium pentanesulfonate/L of the solution, adjusted to pH 3.0 with orthophosphoric acid. The flow rate was 1.0 mL/min. The detection was carried out at 215 nm using a diode array detector. The developed method was validated according to the International Conference on Harmonization (ICH) guidelines for specificity, limit of detection, limit of quantification, linearity, precision and robustness. The proposed method is convenient and reliable for the purity control in both raw materials and dosage forms.