Concepedia

Publication | Open Access

Evaluation of a lifetime‐based optode to measure oxygen in aquatic systems

261

Citations

34

References

2006

Year

Abstract

In this article, we evaluate the performance of a commercially available lifetime‐based optode and compare it with data obtained by other methods. We performed a set of 10 different tests, including targeted laboratory evaluations and field studies, covering a wide range of situations from shallow coastal waters and wastewater treatment plants to abyssal depths. Our principal conclusion is that, owing to high accuracy (± 2 µM), long‐term stability (more than 20 months), lack of pressure hysteresis, and limited cross‐sensitivity, this method is overall more suitable for oxygen monitoring than other methods.

References

YearCitations