Catheterization and Cardiovascular Diagnosis · 1990 · 30 citations · 5 references
MedicineCardiologySurgeryCardiothoracic SurgeryInterventional CardiologyBalloon AngioplastyCardiac CatheterizationCatheter ManagementCoronary AngioplastyCoronary Artery SurgeryPercutaneous Coronary InterventionAngioplasty ComplicationsBalloon Catheter DilationBalloon Catheter Fragment RetrievalBalloon RuptureCoronary Balloon AngioplastyEntrapmentRetrieval
We report a case of balloon catheter rupture with subsequent entrapment during percutaneous transluminal coronary angioplasty. The presence of a calcified, distal lesion is believed to have prevented withdrawal of the broken catheter. A nonsurgical retrieval technique, using a second dilation system, was used to free the catheter.
5
Percutaneous Transluminal Coronary Angioplasty in 1985–1986 and 1977–1981
Katherine M. Detre, Richard Holubkov, Sheryl F. Kelsey et al. · New England Journal of Medicine · 1988 · 855 citations
Medicine, Cardiology, Surgery +14
Entrapment of an angioplasty balloon catheter: A case report
Thomas F. Rizzo, Roland Werres, John Ciccone et al. · Catheterization and Cardiovascular Diagnosis · 1988 · 22 citations
Medicine, Cardiology, Surgery +10