Development of hydrogenated amorphous silicon sensors for high energy photon radiotherapy imaging

Larry E. Antonuk, J. Yorkston, J. Boudry, M. J. Longo, J. Jimenez, R. A. Street

IEEE Transactions on Nuclear Science · 1990 · 39 citations · 7 references

Concepts

Abstract

Measurements with a-Si:H n-i-p photodiodes with sensitive areas of approximately 0.6 mm/sup 2/ exposed to a 6 MV radiation therapy treatment beam have been performed. Such photodiodes can be configured into large two-dimensional arrays of addressable sensors suitable for real-time imaging of megavoltage treatment beams as well as other applications such as diagnostic X-ray imaging. Signal sizes per radiation pulse up to 3.9*10/sup 6/ and 69*10/sup 6/ electrons are observed when the sensors are used with cronex (CaWO/sub 4/) and lanex (gadolinium-oxysulfide) intensifying screens, respectively. For the cronex and lanex screens the size of the measured signals can be enhanced by reducing the thickness of the top p-layer, thereby reducing the attenuation of the incident light. However, this is a smaller effect for the lanex screen, which provides substantially more light signal. The variation in cronex signal size with p-layer thickness is consistent with calculations based on known attenuation coefficients.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

7