IEEE Transactions on Nuclear Science · 1990 · 39 citations · 7 references
EngineeringBiomedical EngineeringElectrical EngineeringRadiation DetectionRadiation MeasurementSensorsDetector TechnologyPhotodetectorsRadiation DetectorsRadiation Imaging DetectorsRadiation TherapySemiconductor DetectorsAmorphous SiliconDetector DesignHydrogenated Amorphous SiliconMegavoltage ImagingPhotodiodesPhoton RadiotherapyReal-time ImagingSilicon Photodiodes
Measurements with a-Si:H n-i-p photodiodes with sensitive areas of approximately 0.6 mm/sup 2/ exposed to a 6 MV radiation therapy treatment beam have been performed. Such photodiodes can be configured into large two-dimensional arrays of addressable sensors suitable for real-time imaging of megavoltage treatment beams as well as other applications such as diagnostic X-ray imaging. Signal sizes per radiation pulse up to 3.9*10/sup 6/ and 69*10/sup 6/ electrons are observed when the sensors are used with cronex (CaWO/sub 4/) and lanex (gadolinium-oxysulfide) intensifying screens, respectively. For the cronex and lanex screens the size of the measured signals can be enhanced by reducing the thickness of the top p-layer, thereby reducing the attenuation of the incident light. However, this is a smaller effect for the lanex screen, which provides substantially more light signal. The variation in cronex signal size with p-layer thickness is consistent with calculations based on known attenuation coefficients.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
7