Journal of Bone and Joint Surgery - British Volume · 2009 · 120 citations · 42 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryAnkle PathologyBone GraftingOrthopedicsAnkle ArthritisAnkle ArthroplastyAnkle OrthopedicsAnkle RadiographyAnkle SurgeryPeriprosthetic OsteolysisAnkle Joint ArthroplastyOsteolytic Bone DiseaseOsteolytic Bone LesionsOsteolytic Lesion FormationPeriprosthetic Bone ResorptionRadiographic Detection Of Osteolysis
Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities > 10 mm in diameter considered to be 'marked'. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous coating. Care should be taken with ankle arthroplasty until more is known about the reasons for these severe osteolyses.
42
P. L. R. Wood, Susan Deakin · Journal of Bone and Joint Surgery - British Volume · 2003 · 569 citations
Michael Pyevich, Charles L. Saltzman, John J. Callaghan et al. · Journal of Bone and Joint Surgery · 1998 · 400 citations
The Agility total ankle arthroplasty. Seven to sixteen-year follow-up.
Stephen I. Knecht, Miriam Estin, John J. Callaghan et al. · PubMed · 2004 · 332 citations