Biotechnology and Bioengineering · 1983 · 49 citations · 11 references
EngineeringBiochemistryChemistryAnalytical ChemistryBiosensorsElectrochemistryElectroanalytical ChemistryElectrochemical BiosensorsElectrochemical DetectionEnzymatic BiosensorsEnzyme AssaysEnzyme-based BiosensorsElectrochemical AssaysElectrochemical BiosensingElectrochemical Enzyme BiosensorsElectrochemical Enzyme SensorsElectrochemical SensingElectrode InterfacesOxidase-based BiosensorsResponse Time
The construction and performance of an enzyme electrode is described which specifically detects lower primary aliphatic alcohols in aqueous solutions. The electrode consists of a commercial Clark-type oxygen electrode on which alcohol oxidase (E.C. 1.1.3.13) and catalase were immobilized. The decrease in electrode current is linearly proportional to ethanol concentrations between 1 and 25 ppm. The response of the electrode remains constant during 400 assays over a period of two weeks. The response time is between 1 and 25 min. Assembly of the electrode takes less than 1 h.
11
Microbial Assimilation of Methanol
Hermann Sahm, Fritz Wagner · European Journal of Biochemistry · 1973 · 139 citations
Construction and study of electrodes using crosslinked enzymes
G. Broun · Analytical Chemistry · 1975 · 107 citations