Proceedings of the National Academy of Sciences · 1997 · 335 citations · 15 references
MedicineBiomedical EngineeringRadiologyBiomedical OpticsDiagnostic ImagingDiagnostic RadiologyBreast Cancer ImagingBreast ExaminationBreast ImagingMammography PhysicsMedical ImagingOptical DiagnosticsContrast EnhancementImage Contrast EffectsImage Contrast EvaluationImage Contrast MechanismsMammographyMammography ImagingRaster ScanningReal-time Processing
We present a novel approach to optical mammography and initial clinical results. We have designed and developed a frequency-domain (110-MHz) optical scanner that performs a transillumination raster scan of the female breast in approximately 3 min. The probing light is a dual-wavelength (690 and 810 nm, 10-mW average power), 2-mm-diameter laser beam, and the detection optical fiber is 5 mm in diameter. The ac amplitude and phase data are processed with use of an algorithm that performs edge effect corrections, thereby enhancing image contrast. This contrast enhancement results in a greater tumor detectability compared with simple light intensity images. The optical mammograms are displayed on a computer screen in real time. We present x-ray and optical mammograms from two patients with breast tumors. Our initial clinical results show that the frequency-domain scanner, even at the present stage of development, has the potential to be a useful tool in mammography.
15
Ballistic 2-D Imaging Through Scattering Walls Using an Ultrafast Optical Kerr Gate
L. Wang, P. P. Ho, C. Liu et al. · Science · 1991 · 550 citations
Bruce J. Tromberg, Olivier Coquoz, Joshua B. Fishkin et al. · Philosophical Transactions of the Royal Society B Biological Sciences · 1997 · 266 citations · Full text
Time-resolved transillumination for medical diagnostics
Stefan Andersson‐Engels, O. Jarlman, Roger Berg et al. · Optics Letters · 1990 · 257 citations · Full text