Method development aspects for the quantitation of pharmaceutical compounds in human plasma with a matrix‐assisted laser desorption/ionization source in the multiple reaction monitoring mode
55
Citations
13
References
2007
Year
The present work investigates various method development aspects for the quantitative analysis of pharmaceutical compounds in human plasma using matrix-assisted laser desorption/ionization and multiple reaction monitoring (MALDI-MRM). Talinolol was selected as a model analyte. Liquid-liquid extraction (LLE) and protein precipitation were evaluated regarding sensitivity and throughput for the MALDI-MRM technique and its applicability without and with chromatographic separation. Compared to classical electrospray liquid chromatography/mass spectrometry (LC/ESI-MS) method development, with MALDI-MRM the tuning of the analyte in single MS mode is more challenging due to interfering matrix background ions. An approach is proposed using background subtraction. With LLE and using a 200 microL human plasma aliquot acceptable precision and accuracy could be obtained in the range of 1 to 1000 ng/mL without any LC separation. Approximately 3 s were required for one analysis. A full calibration curve and its quality control samples (20 samples) can be analyzed within 1 min. Combining LC with the MALDI analysis allowed improving the linearity down to 50 pg/mL, while reducing the throughput potential only by two-fold. Matrix effects are still a significant issue with MALDI but can be monitored in a similar way to that used for LC/ESI-MS analysis.