physica status solidi (a) · 2009 · 49 citations · 28 references
EngineeringNuclear EngineeringPhotonicsRadiation DetectionRadiation MeasurementDetector TechnologyGamma-ray DetectionGamma-ray DetectorsScintillation DetectorsScintillator MaterialsCrystal ScintillatorsDetectorsPhotomultiplier DetectorsPhotomultiplier TubesScintillationScintillation DetectionScintillator CrystalsScintillator Detector Resolution
Abstract The performance of Lu 3 Al 5 O 12 :Ce (LuAG:Ce) and Y 3 Al 5 O 12 :Ce (YAG:Ce) scintillators were compared under γ‐ray excitation using photomultiplier tube (PMT) readout. Light yield non‐proportionality and energy resolution were measured with a Photonis XP5500B PMT. The energy resolution, obtained in this work for 662 keV γ‐rays, was 6.7 and 7.0%, respectively, for LuAG:Ce and YAG:Ce detectors. A fast component in the scintillation decay of LuAG:Ce crystal is faster than that of YAG:Ce crystal, whereas the relative intensity of a fast component for YAG:Ce crystal is higher than that of LuAG:Ce crystal. The coincidence time resolution, obtained in this work for 511 keV annihilation quanta, was 583 and 660 ps, respectively, for LuAG:Ce‐ and YAG:Ce‐based detectors in coincidence experiment together with a BaF 2 scintillation detector. Normalized coincidence time resolution was also discussed in terms of a number of photoelectrons and decay time of the light pulse.
28
Scintillation detectors for x-rays
M. Nikl · Measurement Science and Technology · 2006 · 852 citations
Absolute light output of scintillators
M. Moszyński, M. Kapusta, M. R. Mayhugh et al. · IEEE Transactions on Nuclear Science · 1997 · 428 citations