Publication | Open Access
Magnification Error in Digital Radiographs of the Cervical Spine Against Magnetic Resonance Imaging Measurements
22
Citations
9
References
2013
Year
There were magnification errors on the digital plain radiographs, and they were different in each case. Maximum magnification error differences were 30% (C2) and 31% (C5). Based on these finding, clinicians must pay attention to magnification errors when measuring lengths using digital plain radiography.