A high-resolution SPECT system based on a microchannel-plate imager

F. Avraham Dilmanian, David A. Weber, Jeffrey A. Coderre, D.D. Joel, Kuangyu Shi, G.E. Meinken, P. Som, Yumei Tang, Nora D. Volkow, Conway Yee,

IEEE Transactions on Nuclear Science · 1990 · 19 citations · 1 references

Concepts

Abstract

The authors have developed a high-resolution single-photon emission computed tomography (SPECT) system for small-animal studies with /sup 125/I. The SPECT camera uses a Hamamatsu microchannel-plate imager PIAS (photon-counting image acquisition system), a 0.7-mm-thick NaI(Tl) plate, and a lead-glass parallel-hole collimator. The 15-mm-diameter field of view (FOV) of the PIAS tube was enlarged by an optical fiber faceplate taper to provide a 26-mm camera FOV. Camera absolute sensitivity (counts detected/gamma photons emitted) is 2*10/sup -5/ for /sup 125/I. The system provides an in-plane resolution of about 0.7 mm FWHM (full width at half maximum) in the reconstructed image for a 15-mm radius of rotation. SPECT images from phantom studies and from organs in the rat and mouse provide examples of the resolving capacity of the system.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>