Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system

71

Citations

13

References

2008

Year

Abstract

This work investigates the accuracy of skin dose calculations using the Eclipse treatment planning system. Skin dose was measured using micro-MOSFETs for a range of irradiation conditions (open fields, physical wedges, dynamic wedges, different SSDs) for 6MV and 10MV beams, and compared with the mean dose calculated to a 2mm thick "skin" structure for semi-cylindrical phantoms (representative of a neck or breast). Agreement between the calculated and measured skin dose values was better than +/-20% for 95% of all measured points (both 6MV and 10MV x-ray spectra). For a fixed geometry, the TPS correctly calculated relative changes in dose, showing that minimization of skin dose in IMRT will be effective in Eclipse.