Robotic solid phase extraction and high performance liquid chromatographic analysis of ranitidine in serum or plasma
27
Citations
9
References
1992
Year
Solvent ExtractionSeparation ScienceGas ChromatographyBioanalysisAnalytical ChemistryLiquid ChromatographyClinical ChemistryLaboratory MedicineChromatographyBiochemistryChromatographic AnalysisSample PreparationPharmacologyOn-line Hplc SystemMass SpectrometryForensic ToxicologyInternal StandardMedicineSerial ProcessingDrug Analysis
A fully automated assay for the analysis of ranitidine in serum and plasma, with and without an internal standard, was validated. It utilizes robotic solid phase extraction with on-line high performance liquid chromatographic (HPLC) analysis. The ruggedness of the assay was demonstrated over a three-year period. A Zymark Py Technology II robotic system was used for serial processing from initial aspiration of samples from original collection containers, to final direct injection onto the on-line HPLC system. Automated serial processing with on-line analysis provided uniform sample history and increased productivity by freeing the chemist to analyse data and perform other tasks. The solid phase extraction efficiency was 94% throughout the assay range of 10-250 ng/mL. The coefficients of variation for within- and between-day quality control samples ranged from 1 to 6% and 1 to 5%, respectively. Mean accuracy for between-day standards and quality control results ranged from 97 to 102% of the respective theoretical concentrations.