IEEE Transactions on Instrumentation and Measurement · 2001 · 10 citations · 5 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMetrologyCalibrationEquipment CalibrationMeasurement StandardsCalibration Of Measurement DevicesElectrical Instrument CalibrationImpulse AnalysisImpulse Response MeasurementPulse Generator CircuitsPulse GeneratorsSignal Generator CalibrationStandard Errors Of Measurement
Measurements on a pulse generator have been made at the NRC in Canada and the PTB in Germany. Most of the measured peak values (V/sub p/) among the laboratories agree to /spl plusmn/0.5%, the front times (T/sub 1/) to /spl plusmn/2.0%, and the times to half values (T/sub 2/) to /spl plusmn/1.0%. The differences between calculated calibrator outputs and measured averages were found to be 0.13%, 0.63%, and 0.32% for V/sub p/, T/sub 1/, and T/sub 2/, respectively. Contributions of impulse analysis software to the measurement errors are also discussed.