Publication | Open Access
Performance goals and endpoint assessments for clinical trials of femoropopliteal bare nitinol stents in patients with symptomatic peripheral arterial disease
230
Citations
22
References
2007
Year
Based on the PTA performance efficacy rate of 33% derived from industry clinical trial data and the medical literature, and the requirement that the bare nitinol stent 12-month efficacy performance goal be set to equal twice this rate, the patency efficacy goal equals 66%. Additional information is provided on safety and other reporting standards and stent integrity evaluation for bare metal stents.
| Year | Citations | |
|---|---|---|
Page 1
Page 1