Radiology · 1999 · 122 citations · 33 references
MedicineSurgeryInterventional RadiologyMicrosurgeryVascular SurgeryBiliary InterventionsFistula RepairHepatobiliary SurgeryLiver TransplantationStent PlacementBiliary Fistula ManagementMetallic Biliary StentsPercutaneous Biliary InterventionPortosystemic ShuntsRevisionsShunt MalfunctionStent Graft PlacementTips (Transjugular Intrahepatic Portosystemic Shunt)
PURPOSE: To determine whether polytetrafluoroethylene (PTFE) stent-grafts yield longer patency for creation or revision of transjugular intrahepatic portosystemic shunts (TIPS). MATERIALS AND METHODS: Fourteen PTFE-covered Wallstents were placed in 13 patients with TIPS: seven at shunt creation and seven during revision of TIPS with one to five prior thromboses at 1 day to 1 year after initial TIPS formation. In six cases, prior to stent-graft placement persistent biliary-TIPS fistulas were demonstrated despite repeated shunt revisions with additional metallic stents. RESULTS: All but one graft-lined TIPS were widely patent at a mean duration of venographic follow-up of 19 months (median, 17 months; range, 5-32 months). The limiting percentage of stenosis within the grafted shunts was 0%-10%. One patient developed stent-graft thrombosis; the prior biliary-TIPS fistula was seen despite the graft. A second, parallel PTFE-lined transcaval shunt was created in this patient; it was widely patent at 11-month follow-up. In two asymptomatic patients, stenoses developed in the short, nongrafted portions of the outflow hepatic veins. CONCLUSION: PTFE stent-grafts can markedly prolong TIPS patency, potentially reducing the need for shunt follow-up and revision and the risk of recurrent symptoms associated with shunt stenosis or occlusion.
33