PubMed · 1979 · 43 citations · 10 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyAbdominal ImagingComputed TomographyHepatobiliary ImagingLiver AnatomyLiver DiseaseLiver ImagingContrastLiver Volume AssessmentPhoton EmissionSingle-photon Emission
In 22 volunteers without clinical or laboratory evidence of liver disease, liver volume was determined using single-photon emission computed tomography (ECT). This technique provided excellent object contrast between the liver and its surroundings and permitted calculation of liver volume without geometric assumptions about the liver's configuration. Reproducibility of results was satisfactory, with a root-mean-square error of less than 6% between duplicate measurements in 15 individuals. The volume measurements were validated by the use of phantoms.
10
Journal of the American Medical Association · 1957 · 394 citations
Medicine, Anatomy, Physiology +10
Michael E. Phelps · Seminars in Nuclear Medicine · 1977 · 176 citations